Wrist Digital Blood Pressure Monitor Model UB-543 Instruction Manual Original Manuel d’instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione 使用手冊 1WMPD4002838A 翻譯
Contents Dear Customers................................................................................................................................................... 2 Preliminary Remarks ........................................................................................................................................... 2 Precautions..........................................................................................................................................................
Dear Customers Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this device will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using the device for the first time. Preliminary Remarks This device conforms to the European Directive 93/42 EEC for Medical Products. This is made evident by the mark of conformity.
Do not modify the device. It may cause accidents or damage to the device. To measure blood pressure, the wrist must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery. This may cause pain, numbness or a temporary red mark to the wrist. This condition will appear especially when measurement is repeated successively. Any pain, numbness, or red marks will disappear with time. Contraindications The following are precautions for proper use of the device.
Parts Identification button ( average, memory and change of preset parameters) Display SET button ( user selection and clock START ) button Battery cover Carrying case Cuff Alkaline batteries (LR03 or AAA) Display WHO classification indicator and Pressure bar indicator Heart mark User and User MEMORY Average Battery indicator Systolic pressure Diastolic pressure Correct Position Guidance (C.P.G.) indicator Irregular Heartbeat (I.H.B.
Symbols Symbols that are printed on the device case Symbols SN IP Function / Meaning Standby and Turn the device on Battery installation guide Direct current Serial number Date of manufacture Type BF: Device and cuff are designed to provide special protection against electrical shocks.
Symbols Symbols that appear on the display (continued) Symbols Function / Meaning Previous measurements stored in MEMORY Average data Full Battery The battery power indicator during measurement Low Battery The battery is low when it blinks or Unstable blood pressure due to movement during measurement Recommended Action Replace all batteries with new ones when the indicator blinks. Take another measurement. Remain still during measurement.
Using the Monitor 1. Installing / Changing the Batteries 1. Remove the battery cover. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) are correct. Use only LR03 or AAA batteries. Battery cover 3. Attach the battery cover. Used batteries New batteries Insert the batteries as shown in the battery compartment. If installed incorrectly, the device will not work.
2. Selecting a User 1. Press the Using the Monitor SET button when turning off the device. The indicator or Turn off is blinking. SET button SET button. 2. Select a user from user and user using the Press the START button to turn off the device. After three minutes of non-operation, the device will turn off automatically. User or user START button 3. Adjusting the Built-in Clock Before Use 1. Press and hold the SET button until the year starts blinking. 2. Select the year using the button.
4. Applying the Cuff Using the Monitor 1. Wrap the cuff around your wrist about 1 cm above your hand as shown in the figure at the right. About 1 cm 2. Fasten the cuff tightly using the Velcro strip. Note: For proper measurements, fasten the cuff tightly and measure on a bare wrist. 5. How to Take Proper Measurements For the most accurate blood pressure measurement: Remain still and keep quiet during measurement. Sit down in a comfortable position. Place your elbow on a table.
6. The C.P.G. Function Switch Refer to page 16 for the C.P.G. function that will indicate the proper angle so that the height of the cuff is the same level as your heart. 1. After step 6 in page 8, press the button to select either " " or " " concerning the C.P.G. function. 2. Press the SET button to store the selection. 3. If you do not use my C.P.G. function, press START button to turn off. Proceed to "8. Measurement". If you use my C.P.G. function, proceed to "7. Selecting the C.P.G. and my C.P.G.".
Note: When removing the batteries, preset parameters (of clock, user and my C.P.G.) are reset. SET button. Select a user from user and user using the 8. Measurement During measurement, it is normal for the cuff to feel very tight. 9. After Measurement While readings are displayed, if you press the START button to turn the device off, new readings are stored in memory. button to turn the device off, new readings are not stored.
Measurements Note: The UB-543, once used, will provide an inflation appropriate to the user. 1. Wrap the cuff around your wrist. Sit comfortably with the cuff at the same level as your heart and relax. 2. Press the START button. All of the display segments are displayed. 3. Select a user from user and user using the Adjust and keep the height of the cuff (with blue light) to the same level as your heart using the C.P.G. indicator.
Recalling the Memory Data Turn off Note: This device stores the last 60 measurements in memory. 1. Press the button when turning the device off. The average of all measurements and the number of data are displayed. or START button to turn the If no data, “0” is displayed. Press the device off. 2. Use the following buttons to display data (of number and measurement data). SET button.
Deleting Data Stored in Memory 1. Select a user from user and user device using the START button. using the SET button. Turn off the 2. Press and hold the button until truing off the device automatically. mark, deletes data stored in The device displays a user icon and the mark and turns off automatically. memory while blinking the Note: This operation will delete specified user data stored in memory. You cannot select which data to delete.
What is an Irregular Heartbeat? Model UB-543 blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you be relaxed, remain still and do not talk during measurements. symbol frequently.
The C.P.G. Indicator The C.P.G. Indicator The C.P.G. indicator The C.P.G. (Correct Position Guidance) indicator is the function to inform a difference between the height (wrist angle) of the blood pressure monitor and your cardiac height in the correct posture (Example: sitting posture, height of table and chair, etc.) during the measurement. The indicator can be used to get more stable measurement condition. The C.P.G. Indicator The height of the blood pressure monitor is lower than your heart.
What is Blood Pressure? About Blood Pressure Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
WHO Blood Pressure Classification Standards to assess high blood pressure, without regard to age, have been established by the World Health Organization (WHO), as shown in the chart at the right. Blood Pressure Variations An individual’s blood pressure varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals variations are even more pronounced.
Troubleshooting Problem Possible Reason Batteries are drained. Nothing appears in the display, even when the Battery terminals are not in the correct device is turned on. position. The cuff does not inflate. Battery voltage is too low. (LOW BATTERY mark) blinks. If the batteries are drained completely, the mark does not appear. The cuff is not applied properly. You moved your wrist or body during the measurement. The device does not measure. Readings are too high The cuff position is not correct.
Technical Data Type Measurement method Measurement range Measurement accuracy Power supply Number of measurements Wrist circumference Classification Applied part Useful life Clinical test EMC Memory Operating conditions Transport / storage conditions Dimensions Weight Ingress protection UB-543 Oscillometric measurement Pressure: 0 – 299 mmHg Pulse: 40 – 180 beats / minute Pressure: ±3 mmHg Pulse: ±5% 2 x 1.5V alkaline batteries (LR03 or AAA) Approx.
1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LIMITED Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131 U.S.A.