Instruction manual

10. Connect channel #1 to trigger #3.
Press keys: CONNECT,1,3,ENTER (see Section 9).
Channel #1 will now trigger both channels #2 & #3.
Note:
The ‘
sign on the panel above the ‘1.4’ digits. This arrow indicates the direction of the connection.
To demonstrate Master-8’s most powerful feature
Assume that you programmed all the 8 channels to run in a specific pattern (your present experiment)
and now you want to switch to a new pattern (PARADIGM) that also uses all the 8 channels. With
Master-8 you can do this very easily. But first you have to program the other paradigms.
11. Press key: ‘ALL
The name of the present paradigm (‘1’) is displayed.
In order to switch to paradigm #5,
Press keys: ‘5,ENTER (see Section 10).
Note:
Since this is a new paradigm all the channels are off, all the internal connections are
disconnected and all the time parameters have their default values.
12. Set all the 8 channels to the FREE-RUN mode
Press keys: ‘FREE,1,ENTER’, ‘FREE,2,ENTER’. Etc.
Each of the 8 channels now runs continuously at its own rate, independently of the other channels.
13. Now you can switch back to the previous paradigm by recalling it.
Press keys: ‘ALL,1,ENTER’ and you obtain the previous paradigm.
Press keys: ‘ALL,5,ENTER’ and you obtain the other one. All the channels are now working in
synchronization since you started them together.
14. Turn the power off.
Notes:
a. Master-8 stores all the 8 paradigms even when the power is turned off.
b. There is a lithium battery for memory retention. (This battery need only be replaced