Instruction manual
Note:
Only integer numbers are accepted for M. The exponent should equal 1 only for M > 9999.
8.3 Counting the Pulses
Whenever channel 8 is in the TRAIN mode, the display shows how many pulses still remain in the
existing TRAIN.
Press keys: ‘TRAIN,8,ENTER’.
Now trigger channel 8 manually
Press key: ‘8’.
Table 1. Parameters and Error Indication
Parameter Min Max Error indication for illegal
values.
(x is the channel number)
M 1 59,900 Mx Err
DURATION
40µsec
3999 sec Dx Err
DELAY
100µsec
Delay> Duration/10,000
3999 sec Lx
Lx
Err
Err
INTERVAL
a)Normal usage:
b)Train Mode:
c)If the channel
internally activates
other channels:
60µsec
Interval> Duration + 9µsec
Interval> Duration + 59µsec
Interval >500µsec
3999 sec Ix
Rx
Tx
Cx
Err
Err (Rate err)
Err (Train err)
Err (Connect err)
Example: The message ‘R7 Err’ indicates ‘Rate error’ on channel #7. The reason is that the channel is
now running in FREE-RUN, GATE or TRAIN mode and the interval is shorter than the duration or at