A2Z Ozone Systems Inc. BED STERILIZER MANUAL A2Z Ozone Systems Inc. 1844 Cargo Court Louisville, KY 40299 +1 502-499-4977 +1 502-499-4976 fax www.A2Zozone.
Content 1. Learn about the machine before use Before use In use 2. Installing main machine and accessories 3. External structural diagram 4. Operation directions Manual selection Mode auto selection 5. Maintenance and Malfunction 6. Notes 7. Technical instruction 8.
Learn about the machine before use Before installation Forbid placing the Bed Sterilizer upside down or horizontally during transport. Use 220V AC power. Add stabilizing devices before use if the instability of power voltage exceeds ±10%. Insert the plugs of power cords into safe and qualified sockets solidly. Choose the products passing safe electric certification if extended wiring board is needed.
Installing main machine and accessories Shipping list No.
External structural diagram 1. Handle 2. Panel 3. Cabinet Cover 4. Ear Screw 5. Foot wheel 6. Safety foot 7. Aluminum accessories 8. Main Switch 9. Hose 10.
OPERATION DIRECTION Put the mattresses, bed clothes, bed sheets, pillows, etc. into the sealed plastic sterilization bags and then connect the ozone gas transport pipes. Turn on the power button ON. Then choose the sterilization time. Last, press the thin film switch ON, and then the equipment would begin to work. Count down the time, and when the clock reaches 00, the disinfecting operation has finished. After finishing the sterilizing operation, the equipment would ring.
switch “ON”. It will enter into the whole sterilization procedure. The progressing time of each stage is Default. The Default as following: 1. 1 channel is switched on Vacuum: 5 min Ozone generating: 18 min Ozone-maintain: 20 min Ozone decomposition: 6 min 2. 2 channels are switched on Vacuum: 8 min Ozone generating: 25 min Ozone-maintain: 20 min Ozone decomposition: 9 min 3.
After switched “ON”, it shows TIME and SIGN of each stage by turns. The procedure and sign of each stage is as following: Vacuum (CC) Ozone generating (O3) Ozone-maintain (OF)Ozone decomposition (O2) OVER (OP). The procedure could be stopped when vacuum by switch “OFF”, otherwise, in other stages, any press on thin film will be no function. It shows “OP” when whole procedure is completed and buzz 3 times at the same time. After that turn off the main switch.
enter into the sterilization procedure. The Default time is as following: 1. Modo 1 Vacuum: 10 min Ozone generating: 30 min Ozone-maintain: 10 min Ozone decomposition: 20 min The whole sterilization time is 70 min. 2. Modo 2 Vacuum: Ozone generating: Ozone-maintain: Ozone decomposition: 10 min 25 min 10 min 15 min The whole sterilization time is 60 min. NOTES: After start, it shows TIME and SIGN of each stage by turns.
Maintenance Clean the cabinet after used a period of time. Don’t put wastes or water or oil containers on the Bed Sterilizer. Clean the cooling fan at the bottom of device. Handle with care the device. Check for potential gas leakage on hoses and the sealed bag. If there is a leakage, change the bag.
How do you handle problems during use of Bed Sterilizer? Malfunction Reason Suggestions 1. Check if the power plug to be used is the correct voltage No electricity after main No power supply enter for the unit. 2. Check if the socket has switch is on into device electricity. 3. Check the fuse. 1. Turn off the main switch, and restart it after 1 minute. 1. Could not be startup Contact manufacture if it’s not the unit after power is on working. 2.Alarm rings Chipset malfunction 2.
Hospitals Keeping the hospital environment infection free is a prime concern of any conscientious hospital management. Preventing contamination and spread of infection is an assigned responsibility of the doctors and nurses, which can never be compromised. The most common cause of high internal rate of infection in a hospital is the poor sanitization of the beddings (sheets, pillows, mattresses etc.) and the clothing used by the patients.
11.It is unique that can sterilize 3 beds all at the same time, which has increase the efficiency of bedding sterilization, and solve out the problem of bed shortage in hospital obviously. 12. It can apply to the clothing sterilization of staff in operating rooms other than bedding sterilization. 13. It is unique that has total digital command and automatic function.
A2Z Ozone – 90 days LIMITED WARRANTY The limited warranty set forth below applies to products manufactured by A2Z Ozone Systems Inc. - and sold by A2Z Ozone Systems or its authorized dealers. This limited warranty is given only to the first retail purchaser of such products and is not transferable to any subsequent owners or purchasers of such products.
A2Z Ozone WARRANTY REGISTRATION CARD This information can be also send by EMAIL to us. Please provide ALL pertinent information and send within 20 days from purchase to: service@a2zozone.com Otherwise, please fill out completely and return to A2Z Ozone by post within twenty (20) days of purchase.
A2Z Ozone Systems Inc. 1844 Cargo Court Louisville, KY 40299, USA +1 502-499-4977 +1 502-499-4976 fax www.A2Zozone.