Use Manual

Table Of Contents
Error with pipetting
Probable cause Corrective action
The wrong sample tube size was used. Only use sample tubes that are approved for the
laboratory automation system.
Related information...
Errors and troubleshooting, page 36
CAR with sample at AccessPoint, in traffic jam, or in long queue;
sample matter has not been removed
Probable cause Corrective action
The analyzer is not ready for operation or the laboratory
automation system connection was disrupted.
1. Transition the analyzer to an operable status.
2. Observe the manufacturer's instructions.
Related information...
Errors and troubleshooting, page 36
CAR with sample does not move to the analyzer
Probable cause Corrective action
An error or defect occurred that involves the switch. Contact an Abbott Laboratories representative or an
authorized service representative.
An analyzer error occurred. Resolve the error on the analyzer.
An error occurred with the Track Sample Manager
(TSM) or the Track Workflow Manager (TWM).
1. Verify the TSM or TWM connection.
2. Contact an Abbott Laboratories representative or
an authorized service representative if necessary.
Related information...
Errors and troubleshooting, page 36
Transport ceased on the track
Probable cause Corrective action
A fuse failure occurred. Contact an Abbott Laboratories representative or an
authorized service representative.
Related information...
Errors and troubleshooting, page 36
Section 9
Troubleshooting
Errors and troubleshooting
Sample Access Line (SAL) Interface Manual
80003970-101 DRAFT - 2022-MM-DD
37