Use Manual
Table Of Contents
- Table of contents
- Read me first
- Section 1: Use or function
- Section 2: Installation procedures and special requirements
- Section 3: Principles of operation
- Section 4: Performance characteristics and specifications
- Section 5: Operating instructions
- Section 6: Operational precautions and limitations
- Section 7: Hazards
- Section 8: Service, maintenance, and diagnostics
- Section 9: Troubleshooting
- Appendix A: Alinity ci-series SAL
- Design and function
- Technical data
- Sample Access Line operation
- Buttons
- Start screen
- Power on the Sample Access Line
- Place the Sample Access Line online
- Place the Sample Access Line offline
- Pause the Sample Access Line
- Deactivate pause mode
- Restart the Sample Access Line
- Main menu screen element descriptions
- Login screen
- Access the Login screen
- Information screen element descriptions
- Access the Information screen
- Access the Configuration screen
- Error message screen
- Acknowledge an error message
- Acknowledge a message
- Weekly cleaning procedure
- Error messages
- Communication and software update error messages (0-10000)
- Module component error messages (20000-29999)
- Message code: 20105
- Message code: 20200
- Message code: 20201
- Message code: 20202
- Message code: 20203
- Message code: 20204
- Message code: 20205
- Message code: 20206
- Message code: 20207
- Message code: 20209
- Message code: 20210
- Message code: 20211
- Message code: 20901
- Message code: 20902
- Message code: 20910
- Message code: 20911
- Message code: 20912
- Archive error messages (21300-21468)
- Message code: 22501
- Message code: 22502
- Message code: 22550
- Message code: 22600
- Message code: 22601
- Message code: 22602
- Message code: 22603
- Message code: 22604
- Message code: 22605
- Message code: 22606
- Message code: 22607
- Message code: 22608
- Display error messages (23000-23017)
- Message code: 27000
- Error message on the touchscreen user interface
- CAR with sample at AccessPoint, in traffic jam, or in long queue; sample matter has been removed
- Revision history
- Index
Error with pipetting
Probable cause Corrective action
The wrong sample tube size was used. Only use sample tubes that are approved for the
laboratory automation system.
Related information...
Errors and troubleshooting, page 36
CAR with sample at AccessPoint, in traffic jam, or in long queue;
sample matter has not been removed
Probable cause Corrective action
The analyzer is not ready for operation or the laboratory
automation system connection was disrupted.
1. Transition the analyzer to an operable status.
2. Observe the manufacturer's instructions.
Related information...
Errors and troubleshooting, page 36
CAR with sample does not move to the analyzer
Probable cause Corrective action
An error or defect occurred that involves the switch. Contact an Abbott Laboratories representative or an
authorized service representative.
An analyzer error occurred. Resolve the error on the analyzer.
An error occurred with the Track Sample Manager
(TSM) or the Track Workflow Manager (TWM).
1. Verify the TSM or TWM connection.
2. Contact an Abbott Laboratories representative or
an authorized service representative if necessary.
Related information...
Errors and troubleshooting, page 36
Transport ceased on the track
Probable cause Corrective action
A fuse failure occurred. Contact an Abbott Laboratories representative or an
authorized service representative.
Related information...
Errors and troubleshooting, page 36
Section 9
Troubleshooting
Errors and troubleshooting