Technical data
NOTE: Only sample tubes approved for the laboratory automation system (LAS) may be used.
NOTE: Throughput depends on the Alinity ci-series configuration and the input of samples on
the GLP systems Track.
NOTE: The safety instructions in the GLP systems Track Operations Manual must be observed.
Table 1: Technical data
Throughput 200 samples per hour for each connected Alinity ci-
series analyzer
Table 2: Environmental specifications
Altitude 30.8 m (100 ft) below sea level to 2000 m (6561 ft)
above sea level
Ambient temperature during operation 15°C to 30°C
Relative humidity during operation 30% to 80%, noncondensing
Acoustic level:
SAL
Maximum 65 dBA
Alinity ci-series
Comply with the manufacturer specifications.
NOTE: Electrical lines for the SAL are routed in the power ducts of the GLP systems Track. The
outlets installed in the SAL are used only for the LAS.
Table 3: Electrical safety parameters
Installation category II (overvoltage category)
Pollution degree 2
NOTE: Electrical safety parameters have no bearing
on performance.
Table 4: Energy data specifications
Supply voltage 220 V - 230 V
Supply frequency 50 Hz/60 Hz