CAUTION: Restarting the Sample Access Line (SAL) while the analyzer is processing
samples presents a health risk to patients and test results. Ensure that all aspirations
have been completed before restarting the SAL. During initialization, CARs may be
released from AccessPoints and can potentially cause incorrect aspirations or spilled
samples.
Related information...
Sample Access Line operation, page 46
Place the Sample Access Line online
Prerequisite The arrow area of the Online/Offline button is gray.
Required module
status
Offline
Perform this procedure to place the Sample Access Line (SAL) online.
NOTE: The SAL is automatically online after being powered on.
1.
On the main menu, tap the gray area of the Online/Offline button .
2. Wait for the module to transition to a status of Online.
The Online/Offline button and the signal lamp are illuminated steady green. The module is
online.
Related information...
Sample Access Line operation, page 46
Place the Sample Access Line offline
Prerequisite The Online/Offline button is illuminated steady green.
Required module
status
Online
Perform this procedure to place the Sample Access Line (SAL) offline. All processes running in
the module stop. CARs are no longer routed to the module.
NOTE: The SAL is automatically online after being powered on.
1.
On the main menu, tap the green arrow area of the Online/Offline button .
2. Wait for the module to transition to a status of Offline.
The Online/Offline button is gray. The signal lamp is illuminated steady yellow. The module is
offline.