Restart the Sample Access Line
Perform this procedure to restart the module if the user interface (UI) is no longer responding.
1. Before restarting the module, ensure that all analyses have been completed.
CAUTION: Restarting the Sample Access Line (SAL) while the analyzer is processing
samples presents a health risk to patients and test results. Ensure that all
aspirations have been completed before restarting the SAL. During initialization,
CARs may be released from AccessPoints and can potentially cause incorrect
aspirations or spilled samples.
2. At the rear of the UI, press the Reset button [1].
Figure 4: Reset button
Related information...
Sample Access Line operation, page 46
Appendix A
Alinity ci-series SAL
Sample Access Line operation
Sample Access Line (SAL) Interface Manual