Use Manual
Table Of Contents
- Table of contents
- Read me first
- Section 1: Use or function
- Section 2: Installation procedures and special requirements
- Section 3: Principles of operation
- Section 4: Performance characteristics and specifications
- Section 5: Operating instructions
- Section 6: Operational precautions and limitations
- Section 7: Hazards
- Section 8: Service, maintenance, and diagnostics
- Section 9: Troubleshooting
- Appendix A: Alinity ci-series SAL
- Design and function
- Technical data
- Sample Access Line operation
- Buttons
- Start screen
- Power on the Sample Access Line
- Place the Sample Access Line online
- Place the Sample Access Line offline
- Pause the Sample Access Line
- Deactivate pause mode
- Restart the Sample Access Line
- Main menu screen element descriptions
- Login screen
- Access the Login screen
- Information screen element descriptions
- Access the Information screen
- Access the Configuration screen
- Error message screen
- Acknowledge an error message
- Acknowledge a message
- Weekly cleaning procedure
- Error messages
- Communication and software update error messages (0-10000)
- Module component error messages (20000-29999)
- Message code: 20105
- Message code: 20200
- Message code: 20201
- Message code: 20202
- Message code: 20203
- Message code: 20204
- Message code: 20205
- Message code: 20206
- Message code: 20207
- Message code: 20209
- Message code: 20210
- Message code: 20211
- Message code: 20901
- Message code: 20902
- Message code: 20910
- Message code: 20911
- Message code: 20912
- Archive error messages (21300-21468)
- Message code: 22501
- Message code: 22502
- Message code: 22550
- Message code: 22600
- Message code: 22601
- Message code: 22602
- Message code: 22603
- Message code: 22604
- Message code: 22605
- Message code: 22606
- Message code: 22607
- Message code: 22608
- Display error messages (23000-23017)
- Message code: 27000
- Error message on the touchscreen user interface
- CAR with sample at AccessPoint, in traffic jam, or in long queue; sample matter has been removed
- Revision history
- Index
Related information...
Module component error messages (20000-29999), page 62
Message code: 20902
Invalid reference positions!
Probable cause Corrective action
An error has occurred. 1. Correct the error by following the instructions on
the touchscreen user interface.
2. If the error cannot be corrected, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
Module component error messages (20000-29999), page 62
Message code: 20910
Invalid JSON checksum.
Probable cause Corrective action
An error has occurred. 1. Correct the error by following the instructions on
the touchscreen user interface.
2. If the error cannot be corrected, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
Module component error messages (20000-29999), page 62
Message code: 20911
Inplausible reference position!
Probable cause Corrective action
An error has occurred. 1. Correct the error by following the instructions on
the touchscreen user interface.
2. If the error cannot be corrected, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
Module component error messages (20000-29999), page 62
Message code: 20912
A component has a wrong fw version.
Appendix A
Alinity ci-series SAL
Error messages
Sample Access Line (SAL) Interface Manual
80003970-101 DRAFT - 2022-MM-DD
67