4. Load the sample tubes in a straight, upright position.
5. Be sure to seat the sample tubes correctly in the FlexRacks.
6.
To close the drawer, press the push button above the drawer (or tap the Open/Close button
on the touchscreen user interface).
Related information...
Input/Output Module operation, page 162
Sample tube specifications, page 83
Sample bar code label requirements, page 83
Replace a FlexRack (IOM), page 308
Unload samples from FlexRacks (IOM)
Required module
status
Online
Perform this procedure to unload sample tubes from FlexRacks.
CAUTION: Biological RISKS. This activity or area may expose you to potentially
infectious material.
CAUTION: DO NOT REACH INTO DRAWERS. Injury or bodily harm can occur when
accessing the module interior while the module is powered on. Do not access the
drawers while they are opening or closing. Only load the samples when the drawers are
fully open. The drawer push button blinks when it is unsafe to access the drawer. Never
reach into the module through the drawers.
1.
To open the appropriate drawer, press the push button above the drawer (or tap the Open/
Close button on the touchscreen user interface).
IMPORTANT: Do not pull on the drawer handle to open the drawer. Pulling on the drawer
handle causes an error status for the drawer and does not fully open the drawer.
2. Remove the sample tubes from the FlexRack.
3.
To close the drawer, press the push button above the drawer (or tap the Open/Close button
on the touchscreen user interface).
Related information...
Input/Output Module operation, page 162
Replace a FlexRack (IOM), page 308