6. Press the drawer push button to close the drawer.
Related information...
As-needed cleaning procedures (IOM), page 201
Clean the device-specific racks of the analyzer (IOM)
Required module
status
Offline
Perform this as-needed procedure to clean the device-specific racks of the analyzer.
1. Ensure that all samples are removed from the module.
2. See the manufacturer specifications for the device-specific rack.
Related information...
As-needed cleaning procedures (IOM), page 201
Clean the AccessPoint (IOM)
Required materials
Laboratory-grade surface disinfectant
Lint-free cloth
Required module
status
Offline
Perform this as-needed procedure to clean the AccessPoint.
1. Ensure that all samples are removed from the module.
2. Open the front module cover.
3. Dampen a lint-free cloth with a surface disinfectant.
4. Carefully wipe the AccessPoint to remove any dust.
5. Close the front module cover.
Related information...
As-needed cleaning procedures (IOM), page 201
Clean the bar code readers (IOM)
Required materials
Gentle, antistatic glass cleaner
Lint-free cloth
Required module
status
Offline
Service, maintenance, and diagnostics
Input/Output Module maintenance
Section 9
204
GLP systems Track Operations Manual
80003968-101 - 2021-09-16