28128, page 251
Message code: 28100
Sample ... (barcode …) expected on CAR ... but not found.
Probable cause Corrective action
An error has occurred. Search for sample, locate the last known position to
recover. If not resolved, contact Abbott personnel.
Related information...
Controller interfaces sent by module controllers (28000-30100), page 248
Message code: 28104
Sample entry: Barcode verification failed for sample ... (barcode ...) at module .... Barcode read: ....
Probable cause Corrective action
An error has occurred. Verify sample barcode placement and and ensure
it meets requirements. Ensure barcode reader is
functioning. If not resolved, contact Abbott service
personnel
Related information...
Controller interfaces sent by module controllers (28000-30100), page 248
Message code: 28105
Sample exit: Barcode verification failed for sample ... (barcode ...) at module .... Barcode read: ....
Probable cause Corrective action
An error has occurred. Verify sample barcode placement and and ensure
it meets requirements. Ensure barcode reader is
functioning. If not resolved, contact Abbott service
personnel
Related information...
Controller interfaces sent by module controllers (28000-30100), page 248
Message code: 28114
Analyzer error ...: Analyzer ... reported error for sample ... (barcode ...).
Probable cause Corrective action
An error has occurred. Check sample and ensure functioning module. If not
resolved, contact Abbott service personnel.
Related information...
Controller interfaces sent by module controllers (28000-30100), page 248
Message code: 28116
Failed to create aliquot ... from primary sample ... on CAR .... Error code: ....
Section 10
Troubleshooting
Error messages
GLP systems Track Operations Manual
80003968-101 - 2021-09-16
249