Use Manual

Table Of Contents
Input/Output Module configuration
Areas in the module drawers can be configured for sample input and output, loading and
unloading device-specific racks, and a manual archive.
Related information...
Installation procedures and special requirements, page 61
Using drawers with FlexRacks (IOM), page 64
Using device-specific racks and the archive function (IOM), page 68
Using drawers with FlexRacks (IOM)
Each FlexRack or each drawer may be subdivided into different areas. In this process, each
area is assigned its designated usage.
Designated usage Description
Error The error area is used to load any samples with errors
automatically. This process requires the operator to remove
the samples manually, troubleshoot the causes of the errors,
and return the samples to the distribution area.
Input Samples manually placed into the input area are placed into
the distribution system.
Output Samples are automatically sorted into the respective, defined
output areas. This process requires the operator to remove
the samples manually.
Related information...
Input/Output Module configuration, page 64
Defining areas (IOM), page 64
Create an area in the module (IOM), page 65
Configure an area in the module (IOM), page 65
Edit properties of an area (IOM), page 66
View information about an area (IOM), page 67
Delete an area (IOM), page 67
Defining areas (IOM)
Input and output areas are defined first by creating the area on the touchscreen user interface.
Then, each area is configured with a unique ID and color.
The Configuration screen displays options to create and configure areas.
Installation procedures and special requirements
Input/Output Module configuration
Section 2
64
GLP systems Track Operations Manual
80003968-101 - 2021-09-16