3. Insert the new rack.
4. Close the drawer.
The New Rack-ID? screen is displayed.
5.
Tap Yes, the RackId has changed.
6.
Tap the Right Arrow button to go to the next screen.
7. On the Enter RackId screen, type the new rack ID.
8.
To save the rack ID and exit the screen, tap the Right Arrow button.
NOTE: If the drawer is opened and closed without replacing the rack, tap the No, the RackID
has not changed. option on the New Rack-ID? screen. Tap the Right Arrow button to exit the
screen and continue using the assigned rack ID.
Related information...
Using device-specific racks and the archive function (IOM), page 68
Use a drawer to load or unload an analyzer or device-specific rack (IOM)
Prerequisite A drawer has been configured in accordance with the service
requirements.
Required module
status
Online
Perform this procedure to use a drawer to load or unload a device-specific rack on the module.
NOTE: The setting to configure RackPorts and their designated usage must be configured by an
Abbott Laboratories representative or an authorized service representative.
1. Open the drawer.
2. Load or unload a device-specific rack on the RackPort.
3. Close the drawer.
Related information...
Using device-specific racks and the archive function (IOM), page 68