Use Manual
Table Of Contents
- Table of contents
- Read me first
- System documentation
- Section 1: Use or function
- Section 2: Installation procedures and special requirements
- Section 3: Principles of operation
- Section 4: Performance characteristics and specifications
- Section 5: Operating instructions
- Start the GLP systems Track
- Shut down the GLP systems Track
- Perform an emergency shutdown
- Log on to the Track Sample Manager (TSM)
- Log on to the Track Workflow Manager (TWM)
- Unlock and lock the track hoods
- Track Sample Manager user interface
- Roles and permissions (TSM)
- Main menu screen (TSM)
- Main menu screen element descriptions (TSM)
- Track / RE tab (TSM)
- Track / RE tab element descriptions (TSM)
- Controller tab element descriptions (TSM)
- Segment Controller screen element descriptions (TSM)
- Module Controller screen element descriptions (TSM)
- Circles tab element descriptions (TSM)
- Layout tab element descriptions (TSM)
- Statistic tab element descriptions (TSM)
- Modify statistical criteria (TSM)
- Firmware tab element descriptions (TSM)
- Routing Engine tab element descriptions (TSM)
- View the connections to TWM on TSM
- Operation mode tab element descriptions (TSM)
- Samples tab (TSM)
- Samples tab element descriptions (TSM)
- Search tab (Samples) element descriptions (TSM)
- Search result screen element descriptions (TSM)
- Sample Detail screen element descriptions (TSM)
- Sample History screen element descriptions (TSM)
- Search for samples (TSM)
- Remove a sample (TSM)
- Expert search tab (Samples) element descriptions (TSM)
- List requested tab (Samples) element descriptions (TSM)
- Request to send a sample to an output area (TSM)
- CARs tab (TSM)
- Track Workflow Manager user interface
- Roles and permissions (TWM)
- Main menu screen (TWM)
- Main menu screen element descriptions (TWM)
- Menu overview screen element descriptions (TWM)
- Sample search screen (TWM)
- Messages screen (TWM)
- Instrument status screen (TWM)
- Test status screen (TWM)
- Test status screen element descriptions (TWM)
- View the test status (TWM)
- Disable the tests by instrument on the Test status screen (TWM)
- Enable the tests by instrument on the Test status screen (TWM)
- Disable the tests by type of analysis on the Test status screen (TWM)
- Enable the tests by type of analysis on the Test status screen (TWM)
- Connections screen (TWM)
- Workflow screen (TWM)
- Master data screen (TWM)
- Admin screen (TWM)
- Input/Output Module operation
- Section 6: Calibration procedures
- Section 7: Operational precautions and limitations
- Section 8: Hazards
- Section 9: Service, maintenance, and diagnostics
- Section 10: Troubleshooting
- Error messages
- Communication and software update error messages (0-10000)
- Track element error messages (10001-19999)
- Module component error messages (20000-29999)
- Message code: 20105
- Message code: 20150
- Message code: 20151
- Message code: 20200
- Message code: 20201
- Message code: 20202
- Message code: 20203
- Message code: 20204
- Message code: 20205
- Message code: 20206
- Message code: 20207
- Message code: 20209
- Message code: 20210
- Message code: 20211
- Message code: 20300
- Message code: 20900
- Message code: 20901
- Message code: 20902
- Message code: 20906
- Message code: 20907
- Message code: 20908
- Message code: 20909
- Message code: 20910
- Message code: 20911
- Message code: 20912
- Robot PICCOLA error messages (21000-21060)
- Message code: 21000
- Message code: 21001
- Message code: 21002
- Message code: 21003
- Message code: 21004
- Message code: 21005
- Message code: 21006
- Message code: 21008
- Message code: 21009
- Message code: 21010
- Message code: 21011
- Message code: 21012
- Message code: 21013
- Message code: 21014
- Message code: 21015
- Message code: 21016
- Message code: 21017
- Message code: 21018
- Message code: 21019
- Message code: 21021
- Message code: 21023
- Message code: 21034
- Message code: 21036
- Message code: 21037
- Message code: 21041
- Message code: 21050
- Message code: 21060
- Display error messages (23000-23017)
- Message code: 23100
- Drawer error messages (25000-25107)
- Message code: 25000
- Message code: 25001
- Message code: 25002
- Message code: 25003
- Message code: 25005
- Message code: 25006
- Message code: 25009
- Message code: 25010
- Message code: 25014
- Message code: 25015
- Message code: 25016
- Message code: 25017
- Message code: 25018
- Message code: 25019
- Message code: 25028
- Message code: 25100
- Message code: 25101
- Message code: 25102
- Message code: 25103
- Message code: 25104
- Message code: 25105
- Message code: 25150
- Message code: 27000
- Laboratory automation system error messages
- Track Sample Manager errors
- Track element errors (700-16000)
- Controller interfaces sent by module controllers (28000-30100)
- TSM generated message error messages (70000-71999)
- Message code: 70001
- Message code: 70002
- Message code: 70003
- Message code: 70004
- Message code: 70005
- Message code: 70006
- Message code: 70007
- Message code: 70008
- Message code: 70009
- Message code: 70011
- Message code: 70015
- Message code: 70016
- Message code: 70017
- Message code: 70018
- Message code: 70019
- Message code: 70021
- Message code: 70022
- Message code: 70023
- Message code: 70026
- Message code: 70029
- Message code: 70031
- Message code: 70032
- Message code: 70033
- Message code: 70034
- Message code: 70035
- Message code: 70036
- Message code: 70037
- Message code: 70038
- Message code: 70039
- Message code: 70040
- Message code: 70041
- Message code: 70042
- Message code: 70043
- Message code: 70044
- Message code: 70045
- Message code: 70047
- Message code: 70048
- Message code: 70050
- Message code: 70051
- Message code: 70052
- Message code: 70053
- Message code: 70054
- Message code: 70062
- Message code: 70063
- Message code: 70074
- Message code: 70075
- Message code: 70076
- Message code: 70077
- Message code: 70078
- Message code: 70079
- Message code: 70080
- Message code: 70086
- Message code: 70087
- Message code: 70088
- Message code: 70089
- Message code: 70090
- Message code: 70091
- Message code: 70094
- Message code: 70096
- Message code: 70097
- Message code: 70098
- Message code: 70099
- Message code: 70100
- Message code: 70101
- Message code: 70102
- Message code: 70103
- Message code: 70104
- Message code: 70105
- Message code: 70107
- Message code: 70108
- Message code: 70109
- Message code: 70112
- Message code: 70113
- Message code: 70114
- Message code: 70115
- Message code: 70116
- Message code: 70117
- Message code: 70118
- Message code: 70119
- Message code: 70120
- Message code: 70121
- Message code: 70122
- Message code: 70123
- Message code: 70124
- Message code: 70125
- Message code: 70126
- Message code: 71000
- Message code: 71001
- Message code: 71002
- Message code: 71003
- Message code: 71004
- Message code: 71005
- Message code: 71006
- Message code: 71007
- Message code: 71008
- Message code: 71009
- Message code: 71010
- Message code: 71011
- Message code: 71012
- Message code: 71013
- Message code: 71014
- Message code: 71015
- Message code: 71016
- Message code: 71017
- Message code: 71018
- Warning messages of controller interfaces sent by analyzers (80000-80999)
- Message code: 80000
- Message code: 80001
- Message code: 80002
- Message code: 80003
- Message code: 80004
- Message code: 80005
- Message code: 80006
- Message code: 80007
- Message code: 81005
- Message code: 81006
- Message code: 81010
- Message code: 81011
- Message code: 81012
- Message code: 81013
- Message code: 81014
- Message code: 81015
- Message code: 81016
- Message code: 81017
- Message code: 81019
- Message code: 81020
- Message code: 81022
- Message code: 81023
- Message code: 81027
- Message code: 81028
- Message code: 81029
- TWM generated message error messages (81000-82999)
- Observed problems
- CAR observed problems
- New CAR does not move
- Sample tube is slanted in the sample holder
- CAR position is skewed on the track
- CAR is caught on the charge lane charging pin or lane element transition
- CAR movement is jerky or noisy
- Rechargeable lithium battery is not charged
- Rechargeable lithium battery is overheating
- CAR lurches from side to side and is not stopped at the AccessPoint
- CAR LEDs blink in an uncontrolled manner (such as red and yellow) while moving
- CAR always moves at a constant speed
- CAR does not exit the charge lane
- CARs come into contact with the charge lane
- CARs do not move
- CARs are present in the charge lane but cannot move
- CARs move too slowly
- CAR does not maintain distance with the CAR ahead of it or pushes it forward
- Track observed problems
- Track Sample Manager observed problems
- Input/Output Module observed problems
- CAR does not move to the module
- Error message is displayed
- CAR moves through the AccessPoint without stopping
- CAR stops at the AccessPoint and then will not move
- Robot does not work
- Gripper loses its grip on the sample tube
- Sample tube is jammed in the gripper
- Sample tube is tilted in the FlexRack
- Bar code is not read
- CAR observed problems
- Miscellaneous corrective action procedures
- Error messages
- Revision history
- Glossary
- accessories
- AccessPoint
- analyzer
- bar code label
- biological hazard
- CAR
- circle
- Clinical and Laboratory Standards Institute
- CLSI
- control system
- diagnostic procedure
- discipline
- FlexRack
- GLP systems Track
- Input/Output Module
- instrument status
- IOM
- laboratory automation system
- laboratory information system
- lane element
- LAS
- LIS
- loading area
- logon (TSM)
- maintenance procedure
- message
- middleware
- module
- module status
- monitor
- password
- pipettor
- printed circuit board
- RackPort
- radio-frequency identification (RFID) tag
- robot
- screen
- system log
- time-out
- track
- Track Sample Manager
- track section
- Track Workflow Manager
- TSM
- TWM
- user name
- window
- Index
Messages table area
Quantity Displays a column when message grouping according to type
is active. The number of messages of the same type is
indicated. Selecting the Count column expands the messages
group, displaying all older messages of the group. Deselecting
the Group field displays all messages, and the Count column
is no longer displayed in the table.
ID Displays the message ID.
Time Displays the time the message was created
Priority Displays the priority of the message:
Green Informational message
Yellow Warning message
Red Error message
Type Displays the message type with number and text.
Text Displays the detailed message description.
Instrument Displays this message when an instrument is linked. A
magnifying glass icon is displayed when the instrument is
enabled or disabled. A mouse-over displays a pop-up window
with the name of the instrument.
Sample Displays this message when a sample is linked. A magnifying
glass icon is displayed if an error message occurs on the Track
Sample Manager. A mouse-over displays a pop-up window
containing the sample ID or bar code.
Order Displays this message when an order is linked. A magnifying
glass icon is displayed if the laboratory information system
fails to process a message. A mouse-over displays a pop-up
window containing the order ID.
Function button
Refresh Refreshes the messages view.
Section 5
Operating instructions
Track Workflow Manager user interface
GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
147