Use Manual

Table Of Contents
5. Tap Yes, the RackId has changed.
6. Tap the Right Arrow button to go to the next screen.
7. On the Enter RackId screen, type the new rack ID.
8. To save the rack ID and exit the screen, tap the Right Arrow button.
NOTE: If the drawer is opened and closed without replacing the rack, tap the No, the RackID has
not changed. option on the New Rack-ID? screen. Tap the Right Arrow button to exit the screen
and continue using the assigned rack ID.
Related information...
Using racks, page 64
Use a drawer to load or unload from racks
Prerequisite A drawer has been configured in accordance with the service
requirements.
Required module
status
Online
Perform this procedure to use a drawer to load or unload a rack in the module.
NOTE: The setting to configure RackPorts and their designated usage must be configured by an
Abbott Laboratories representative or an authorized service representative.
1. Open the drawer.
2. Load or unload a rack on the RackPort.
3. Close the drawer.
Related information...
Using racks, page 64
Section 2
Installation procedures and special requirements
Input/Output Module configuration
GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
65