Remove a sample (TSM)
Perform this procedure to remove a manually removed sample from the Track Sample Manager
(TSM).
CAUTION: Do not remove samples from a CAR or the track. All samples removed
from the track must be placed back in the Input/Output Module (IOM) for appropriate
routing. When samples are removed from the track, they must be deleted from TSM
before they are reloaded.
IMPORTANT: A sample that was manually removed from the track must be marked as Removed.
Consequently, the sample receives a status of Unknown.
1. On the TSM main menu, tap Samples.
2. On the Samples tab, tap the Search tab.
3. On the Search tab, enter the TSM sample number or the sample bar code.
NOTE: The sample bar code can be entered in the following ways:
The bar code 0999.0999-9 can be entered as 0999.09999 if the exact bar code is
known.
The bar code 0999.0999-9 can be entered as 0999.0999* if the last digit of the bar code
is unknown or if there are multiple samples on the track that end with this bar code. All
samples that have a bar code beginning with 0999.0999 are listed.
4. Tap Search or press the Enter key to display a list of search results.
5. If multiple samples are displayed for one bar code, select the sample based on its status.
NOTE: The following actions can cause multiple samples to be displayed for one bar code:
A bar code containing an asterisk was entered.
Tubes have been repeatedly removed from and placed back on the GLP systems Track.
Additional secondary tubes with an identical bar code have been generated on the
Aliquot Module.
6. Tap the sample to be removed.
7. On the Sample Detail screen, tap Remove.
8. When a confirmation message is displayed, tap OK.
In the Error text box on the Sample Detail screen, the sample is marked as removed.
Related information...
Samples tab (TSM), page 118
Expert search tab (Samples) element descriptions (TSM), page 126
Replace a CAR with a sample, page 304