Use Manual

Table Of Contents
Messages table area
Quantity Displays a column when message grouping according to type
is active. The number of messages of the same type is
indicated. Selecting the Count column expands the messages
group, displaying all older messages of the group. Deselecting
the Group field displays all messages, and the Count column
is no longer displayed in the table.
ID Displays the message ID.
Time Displays the time the message was created
Priority Displays the priority of the message:
Green Informational message
Yellow Warning message
Red Error message
Type Displays the message type with number and text.
Text Displays the detailed message description.
Instrument Displays this message when an instrument is linked. A
magnifying glass icon is displayed when the instrument is
enabled or disabled. A mouse-over displays a pop-up window
with the name of the instrument.
Sample Displays this message when a sample is linked. A magnifying
glass icon is displayed if an error message occurs on the Track
Sample Manager. A mouse-over displays a pop-up window
containing the sample ID or bar code.
Order Displays this message when an order is linked. A magnifying
glass icon is displayed if the laboratory information system
fails to process a message. A mouse-over displays a pop-up
window containing the order ID.
Function button
Refresh Refreshes the messages view.
Section 5
Operating instructions
Track Workflow Manager user interface
GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
147
DRAFT 2022-03-11