Use Manual

Table Of Contents
2. On the menu overview, tap Instrument status.
3. In the Dis-/enable tests column on the Instrument status screen, tap the appropriate
instrument for which one or more tests are to be disabled.
4. When the tests of the analyzer or module selected are displayed, tap to clear the blue check
box in the TWM status column and in the rows of the tests to be disabled.
NOTE: To disable all tests of an instrument at one time, tap to clear the appropriate check
box next to the TWM status column header.
5. Tap Save.
The tests are now disabled.
Related information...
Instrument status screen (TWM), page 148
Enable the tests on the Instrument status screen (TWM)
Perform this procedure to enable the tests on the Instrument status screen of the Track
Workflow Manager (TWM). Individual tests of an analyzer or module can be enabled so that
they can be used as route targets.
1. On the TWM main menu, tap Run.
2. On the menu overview, tap Instrument status.
3. In the Dis-/enable tests column on the Instrument status screen, tap the appropriate
instrument for which one or more tests are to be enabled.
4. When the tests of the analyzer or module selected are displayed, tap the white check box in
the TWM status column and in the rows of the tests to be enabled.
NOTE: To enable all tests of an instrument at one time, tap the appropriate check box next
to the TWM status column header.
5. Tap Save.
The tests are now enabled.
Related information...
Instrument status screen (TWM), page 148
Test status screen (TWM)
On the Test status screen, the operator can perform the following functions:
View the test status.
Disable one or more tests by instrument.
Enable one or more tests by instrument.
Disable one or more tests by type of analysis.
Enable one or more tests by type of analysis.
Operating instructions
Track Workflow Manager user interface Section 5
152 GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11