Load samples into FlexRacks
Prerequisite Become familiar with sample specifications and requirements
and sample bar code label requirements.
Required module
status
Online
Perform this procedure to load sample tubes into FlexRacks in the module drawer.
CAUTION: Biological RISKS. This activity or area may expose you to potentially
infectious material.
CAUTION: DO NOT REACH INTO DRAWERS. Injury or bodily harm can occur when
accessing the module interior while the module is powered on. Do not access the
drawers while they are opening or closing. Only load the samples when the drawers are
fully open. The drawer push button blinks when it is unsafe to access the drawer. Never
reach into the module through the drawers.
CAUTION: Incorrectly loaded samples may constitute a health hazard for patients.
The input area can be configured in different ways, which may increase the risk of
samples being loaded incorrectly. Samples loaded in the wrong area can cause delayed
or incorrect results, and can cause highly sensitive analyzers to malfunction. Only allow
trained personnel to operate the laboratory automation system (LAS). Ensure that
samples are loaded into the defined area for the sample type (for example, centrifuged,
uncentrifuged, capped, or uncapped) on the LAS.
Section 5
Operating instructions
Input/Output Module operation
GLP systems Track Operations Manual