Perform this procedure to unload sample tubes from FlexRacks in the module drawer.
CAUTION: Biological RISKS. This activity or area may expose you to potentially
infectious material.
CAUTION: DO NOT REACH INTO DRAWERS. Injury or bodily harm can occur when
accessing the module interior while the module is powered on. Do not access the
drawers while they are opening or closing. Only load the samples when the drawers are
fully open. The drawer push button blinks when it is unsafe to access the drawer. Never
reach into the module through the drawers.
1. To open the appropriate drawer, press the push button above the drawer (or tap the Open/
Close
button on the touchscreen user interface).
IMPORTANT: Do not pull on the drawer handle to open the drawer. Pulling on the drawer
handle causes an error status for the drawer and does not fully open the drawer.
2. Remove the sample tubes from the FlexRack.
3. To close the drawer, press the push button above the drawer (or tap the Open/Close
button on the touchscreen user interface).