Clean the FlexRacks
Required materials
Laboratory-grade surface disinfectant
Lint-free cloth
Warm water
Required module
status
Offline
Perform this as-needed procedure to clean the FlexRacks.
1. Ensure that all samples are removed from the module.
2. Press the drawer push button to open the drawer.
CAUTION: A minimum clearance between the module and the operator is required
to avoid injury when opening the drawer.
3. Dampen a lint-free cloth with a surface disinfectant.
4.
Carefully wipe each FlexRack to remove any dust.
5. Press the drawer push button to close the drawer.
Related information...
As-needed cleaning procedures, page 203
Clean the drawers
Required materials
Handheld vacuum (recommended)
Laboratory-grade surface disinfectant
Lint-free cloth
Required module
status
Offline
Perform this as-needed procedure to clean the drawers.
1. Ensure that all samples are removed from the module.
2. Press the drawer push button to open the drawer.
CAUTION: A minimum clearance between the module and the operator is required
to avoid injury when opening the drawer.
3. Vacuum the drawer.
4.
Dampen a lint-free cloth with a surface disinfectant.
5. Carefully wipe the drawer to remove any dust.
Section 9
Service, maintenance, and diagnostics
Input/Output Module maintenance
GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
205
DRAFT 2022-03-11