Controller interfaces sent by module controllers (28000-30100), page 247
Message code: 28105
Sample exit: Barcode verification failed for sample ... (barcode ...) at module .... Barcode read: ....
Probable cause Corrective action
An error has occurred. 1. Verify that the placement of the sample bar code is
correct and meets requirements.
2. Ensure that the bar code reader is functioning.
3. If the error is not resolved, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
Controller interfaces sent by module controllers (28000-30100), page 247
Message code: 28114
Analyzer error ...: Analyzer ... reported error for sample ... (barcode ...).
Probable cause Corrective action
An error has occurred. 1. Verify that the sample is present.
2.
Ensure that the module is functioning.
3. If the error is not resolved, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
Controller interfaces sent by module controllers (28000-30100), page
247
Message code: 28116
Failed to create aliquot ... from primary sample ... on CAR .... Error code: ....
Probable cause Corrective action
An error has occurred. 1. Verify that the sample is present.
2.
Ensure that the module is functioning.
3. If the error is not resolved, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
Controller interfaces sent by module controllers (28000-30100), page
247
Troubleshooting
Track Sample Manager errors Section 10
248 GLP systems Track Operations Manual