Use Manual

Table Of Contents
Probable cause Corrective action
An error has occurred. Contact an Abbott Laboratories representative or an
authorized service representative.
Related information...
TSM generated message error messages (70000-71999), page 250
Message code: 70036
Areas ... of module ... cannot be removed. Reason: Areas still contain samples.
Probable cause Corrective action
An error has occurred. Contact an Abbott Laboratories representative or an
authorized service representative.
Related information...
TSM generated message error messages (70000-71999), page 250
Message code: 70037
CAR ... has reported the wrong loading state ... time(s) within ... hour(s).
Probable cause Corrective action
An error has occurred. 1. Locate the CAR and remove it from the track for
maintenance.
2. Contact an Abbott Laboratories representative or
an authorized service representative.
Related information...
TSM generated message error messages (70000-71999), page 250
Message code: 70038
A switch has misrouted CAR ... to the maintenance lane.
Probable cause Corrective action
An error has occurred. 1. Remove the CAR from the maintenance lane and
place it back on the track for further processing.
2.
If the error is not resolved, contact an Abbott
Laboratories representative or an authorized
service representative.
Related information...
TSM generated message error messages (70000-71999), page
250
Message code: 70039
CAR ... requires maintenance and is being routed to maintenance lane ....
Section 10
Troubleshooting
Track Sample Manager errors
GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
259