CARs are present in the charge lane but cannot move
Probable cause Corrective action
The charge lane charging pins do not release the CAR. Contact an Abbott Laboratories representative or an
authorized service representative.
Related information...
CAR observed problems, page 289
CARs move too slowly
Probable cause Corrective action
The CAR requires a restart.
The CAR is defective.
The rechargeable lithium battery is almost empty.
1. Switch off the CAR, and then switch it on again.
2. Contact an Abbott Laboratories representative or
an authorized service representative if necessary.
Related information...
CAR observed problems, page 289
CAR does not maintain distance with the CAR ahead of it or pushes it forward
Probable cause Corrective action
The collision sensor is defective or malfunctioning. 1. Switch off the CAR, and then switch it on again.
NOTE: Do not affix extra labels to CARs as the
labels may adversely affect the collision sensors on
the CARs.
2. Contact an Abbott Laboratories representative or
an authorized service representative if necessary.
Related information...
CAR observed problems, page 289
Track observed problems
Track observed problems include problems that occur with the track.
All work not included in this manual may be performed only by an Abbott Laboratories
representative or an authorized service representative.
Related information...
Observed problems, page 289
Sample is missing, page 295
Operator manually loads or unloads CAR, page 295
Troubleshooting
Observed problems Section 10
294 GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11