Replace the gripper fingers
Prerequisite
The robot is in an accessible position.
All samples must be removed from the module to avoid
sample contamination.
Required materials Tx6 Torx screwdriver
Required module
status
Off
Perform this procedure to replace the gripper fingers on the module.
CAUTION: Overhead obstruction. Operators may hit their heads on open module
covers.
Be aware that injury or bodily harm can occur when module covers are opened and
closed.
Protect the head when working on modules with open module covers.
Frequently observe the functionality of the opening mechanism. Regular visual
inspection of the covers is necessary during maintenance to ensure proper
operation.
NOTE: Inspect all four gripper fingers and replace any defective gripper fingers together with
their screws. The gripper fingers can only be used in one position by design. The procedure for
replacing the gripper fingers is identical for all four fingers.
1. At the front lower end of the module cover, insert the unlock key into the unlock
mechanism.
2. Turn the unlock key counterclockwise a quarter turn.
3. Lift open the cover.
4. Loosen both screws [2] on the gripper finger [1] with the Torx screwdriver.
Troubleshooting
Miscellaneous corrective action procedures Section 10
306 GLP systems Track Operations Manual