4. Remove the damaged FlexRack by pulling it up from the RackPort [3].
5. Insert the new FlexRack [1] on the RackPort [3].
NOTE: Magnets [2] are used to secure the new FlexRack [1] on the RackPort [3].
6. To close the drawer, press the push button above the drawer (or tap the Open/Close
button on the monitor).
Related information...
Input/Output Module corrective action procedures, page 305
Load samples into FlexRacks, page 169
Unload samples from FlexRacks, page 171
Section 10
Troubleshooting
Miscellaneous corrective action procedures
GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
309