Use Manual

Table Of Contents
4. When the confirmation message is displayed, tap yes to delete the area.
The area is deleted.
5. When the Delete area message is displayed, tap OK to acknowledge the update
notification.
6. Tap the Exit button.
The Configure Areas screen is displayed.
7. Switch the module online.
Related information...
Using drawers with FlexRacks, page 60
Place the module online, page 165
Place the module offline, page 165
Using racks
Drawer areas in the Input/Output Module can be used to load and unload sample tubes to
store.
Samples can be inserted into the drawers individually in FlexRacks. Drawers can be opened
independently of each other.
Rack IDs can be assigned either automatically by the module or manually by the operator. The
setting to assign rack IDs automatically or manually is configured by an Abbott Laboratories
representative or an authorized service representative.
NOTE: Areas within the RackPorts are configured by an Abbott Laboratories representative.
Related information...
Input/Output Module configuration, page 60
Replace a rack and set the rack ID, page 64
Use a drawer to load or unload from racks, page 65
Replace a rack and set the rack ID
Prerequisite The rack ID handling feature is set to manual.
Perform this procedure to replace a rack and set a rack ID on the module.
1.
Open the drawer.
2. Remove the rack.
3. Insert the new rack.
4. Close the drawer.
The New Rack-ID? screen is displayed.
Installation procedures and special requirements
Input/Output Module configuration Section 2
64 GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
DRAFT 2022-03-11