to infections due to skin contact with the sample
matter.
Additional sample tube types must be clarified with the device manufacturer and Abbott
Laboratories. For questions or use of formats that are not specified in this manual, contact
an Abbott Laboratories representative or an authorized service representative.
Table 8: Greiner Bio-One MiniCollect Microtubes
Approved Diameter: 13 mm
Height: 75 mm
MiniCollect Microtube is firmly connected to
carrier tube (for children) and can be used on
the Input/Output Module.
Not approved MiniCollect Microtube and carrier tube are
separated. For example:
Serum 450533
EDTA 450530
Carrier tube 450417
Otherwise, no processing can be performed on the
laboratory automation system (LAS).
CAUTION: Risk of injury due to incorrect use of sample tubes. Incorrectly used sample
tubes can cause gripper malfunctions. Spilled sample matter can cause contamination
and severe injuries to the operator. Inadequate sample volume in sample tubes causes
delays in diagnosis and treatment of patients.
Only use sample tubes that are approved for the LAS.
Immediately clean any contaminated areas of the LAS by using a laboratory-grade
surface disinfectant.
If a sample is spilled, ensure that the sample volume in the sample tube is adequate
to perform the remaining tests.