Use Manual

Table Of Contents
1. On the main menu, tap the Configuration button.
2. On the Configuration screen, tap Areas.
3. On the Configure Areas screen, tap new.
The step 1 tab of the Properties menu is displayed.
4. In the list on the left side of the Properties menu, tap the color of the area to configure.
The selected area is displayed in the area display.
5. Tap the drawer for the area to be defined.
The selected drawer is displayed in the component display.
6. In the rack start and rack end boxes, tap the Up Arrow and Down Arrow buttons to define
the FlexRack position inside the drawers.
The positions are arranged from 1 to 5 from the bottom to the top of the drawer.
7. In the position start and position end boxes, tap the Up Arrow and Down Arrow buttons to
define the positions inside the FlexRacks.
The positions are arranged from 1 to 25 from the lower right (position 1) to the upper left
(position 25) of the FlexRack.
8. To confirm the settings and move to the step 2 tab, tap Next.
9. In the Characteristics list in the step 2 tab, tap usage to open the usage list.
NOTE: The Characteristics list has information about the tube type, such as capped, priority,
and fluid type.
10. In the usage list, tap the appropriate usage for the area to configure.
11. To confirm the settings and move to the step 3 tab, tap Next.
NOTE: Only an Abbott Laboratories representative or an authorized service representative
may configure additional settings.
12. To save the configuration settings, tap Update.
To cancel the configuration settings, tap Cancel.
13. When the Update area message is displayed, tap OK to acknowledge the update
notification.
The Configure Areas screen is displayed.
14. Switch the module online.
Related information...
Using drawers with FlexRacks, page 60
Edit properties of an area
Required module
status
Offline
Installation procedures and special requirements
Input/Output Module configuration Section 2
62 GLP systems Track Operations Manual
80003968-101 DRAFT - 2022-03-11
DRAFT 2022-03-11