User's Manual
153
154
Steps Do This
Check your work.3. Conrm that ALL functions are working properly when you performed the
companion self-test. Call Customer Care, if any of the functions are NOT working,
or you get a “Companion Memory Error” warning.
Caution:
If your pump audio or vibratory annunciators are not functioning, switch to insulin
injections immediately AND contact your healthcare professional and/or Customer
Care.
6.8 Pair Companion and Pump
Purpose
To pair your Aviator Companion and Aviator Pump to operate as a system.
When
Do this procedure:
When you receive a new companion or pump that’s a replacement for your system.
•
Materials
Your companion fully charged.•
Your pump with battery installed and initialized.•
Assistance from healthcare professional as needed.•
Helpful Information
Pairing establishes an association of your pump with your companion so that the two will only communicate •
with each other.
Your companion and pump should already be paired when you receive your FreeStyle Aviator System package.
•
The RF will automatically turn ON when pairing, if it is NOT already on.•
Procedure:
Steps Do This
Go to the Companion’s 1.
SETUP menu.
Note: The following steps are performed using your companion.
Notice a. Main is highlighted.
Press b. Enter button.
Scroll to highlight c. SETUP.
Press d. Enter button.
Section 6: When You’re Ready – Optional Settings
6.8 Pair Companion and Pump
Section 6: When You’re Ready – Optional Settings
6.8 Pair Companion and Pump