User's Manual

253
254
Other Preferences That are Not Part of the Wizard
Menu Item Screen
Default
Cong Valid Values Increment
REMINDERS Reminders
1 – 6 On/O
O “On, “O N/A
REMINDERS Reminders
1 – 6 Message
Test Test, “Wakeup, “Basal”, “Bolus”, “Exercise”,
“Meeting”, “Pickup, “Snack”, “Meds”
N/A
REMINDERS Disable All Care Comments On “On, “O N/A
REMINDERS On/O Insulin Temp Care
Comments
O “O, “On N/A
REMINDERS On/O Site Change Care
Comments
O “O, “On N/A
REMINDERS Frequency Site Change Care
Comments
2.0 days 2.0, 2.5, 3.0 N/A
REMINDERS On/O Check BG Care
Comment
O “O, “On N/A
C. Device Specications – Companion
Assay Method (FreeStyle Lite Blood Glucose
meter):
Coulometric electrochemical sensor
Automatic shuto: User settable up to two minutes after last user action
Calibration (FreeStyle Lite Blood Glucose meter): Plasma equivalent
Hematocrit (FreeStyle Lite Blood Glucose meter): 15% to 65%
Measurement Units: Either mg/dL or mmol/L
Result Range (FreeStyle Lite Blood Glucose
meter):
20 to 500 mg/dL (1.1 to 27.8 mmol/L)
Sample (FreeStyle Lite Blood Glucose meter): Whole blood, capillary
Sample size (FreeStyle Lite Blood Glucose
meter):
0.3 microlitre (300 nanolitres)
Test time (FreeStyle Lite Blood Glucose meter): Average of 5 seconds
Dimensions: 3.89 X 2.44 x 0.63” (99 X 62 X 16 mm )
Display Type: Dot graphic•
Display Viewable Dimensions: 1.77 x 1.28” (45 X 32.5 mm)
Display Backlight: Yes
Audio/Vibe Alert: Yes
Weight: 2.6 oz (75 grams)
Companion Battery: Lithium polymer, 3.7V 690mAh
Battery Life: 3 days
The expected life of the battery is dependant upon the following usage:
- 5 boluses per day with total daily bolus delivery of 24 Units (U100)
- Basal delivery of 1U/h (U100) 24 hr/day, 7 days/week
- Prime of 20 Units (U100) occurs 3 times per week.
- 5 blood glucose readings per day
- Enter and log 8 events per day
- Data transfer to CoPilot three times per month
Memory: 90 days worth of events. All events can be seen if downloaded to a PC.
Water Resistance: None
Operational Conditions (Temperature, Humidity
and Pressure):
41°F – 104°F (5°C to 40°C) with 10% to 90% relative
humidity 14.7 psia - 10.2 psia (equivalent to an altitude of sea level to
10,000 feet [3048 meters] under standard atmospheric conditions)
Storage Range Shipping container for 1 year at -4 – 113°F, average temperature at 72.5°F
(-20°C to 45°C, average temperature at 22.5°C) with 10% to 90% relative
humidity, non-condensing.
Data Storage Time: The Companion has the ability to store a minimum of 90 days of
events in non-volatile memory for the expected life of the companion.
Section 12: Appendices
Appendix D: Device Specications – Pump
Section 12: Appendices
Appendix D: Device Specications – Pump