G. Accuracy of Flow Rate
This graph shows the ow rate during the stabilization period. Measurements were made at a basal rate of 1U/h, in
accordance to IEC 60601-2-24:1998 using an Aviator Insulin Reservoir and a Unomedical Comfort infusion set.
0.020
0.015
0.010
0.005
0.000
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400
Time (minutes)
Flow (mL/h)
This graph show the maximum errors observed for specic observation intervals. Measurements were made at
a basal rate of 1U/h, in accordance to IEC 60601-2-24:1998 using an Aviator Insulin Reservoir and a Unomedical
Comfort infusion set.
20
15
10
5
-5
-10
-15
-20
0
15
60 150
Observation Interval (minutes)
330 570 930
Percentage Error of Flow (%)
Overall percentage error: -1.1%
At the minimum rate of 0.05 U/h (0.0005 mL/h), setup errors associated with the IEC 60602-2-24 test are too large
to accurately determine the ow rate of the pump. An alternate test method was used which determines delivery
accuracy by analyzing the movement of drive motor, as reported by an encoder on the motor shaft. Actual insulin
ow rate may be less accurate.
This graph shows the calculated ow rate during the stabilization period. Encoder data was collected at a basal rate
of 0.05U/h.
0.010
0.009
0.006
0.007
0.008
0.003
0.004
0.005
0.002
0.000
0
0.001
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400
Time (minutes)
Flow (mL/h)
This graph show the maximum errors calculated for specic observation intervals. Encoder data was collected at a
basal rate of 0.05U/h.
20
15
10
5
-5
-10
-15
-20
0
15
60 150
Observation Interval (minutes)
330 570 930
Percentage Error of Flow Rate (%)
Overall percentage error: -0.01%