User's Manual

Clinical Trial Support Line: 1-888-971-4766 - 152 -
ART05005 REV 4
To acknowledge alarms:
Press the Right Option Button to turn off a vibration or audio alarm.
Press the Left Option Button to turn off a text display message.
Table 7.2 provides a complete listing of Continuous Monitoring Mode alarm
messages.
Table 7.2: Continuous Monitoring Mode Alarm Messages
Alarm
Message
Urgency* What It Means What To Do
Calibration
Alarms
Do BG Test M A Blood Glucose Mode
test is needed for
calibration. The Receiver
will prompt you to perform
this test approximately 1,
3, and 24 hours after a
new Sensor is inserted.
Perform a Blood Glucose
Mode test (see page 63).
Cal Failed:
Receiver
Temp
Redo BG
Later
I Calibration failed because
the Receiver was too
warm or too cold.
Bring the Receiver to room
temperature, and then
perform another Blood
Glucose Mode test. If
calibration fails 3 times,
remove the Sensor and
insert a new one (see page
43).
Cal Failed
Redo BG
Test
I Calibration result was very
different from the previous
calibration. Another Blood
Glucose Mode test is
needed.
Perform another Blood
Glucose Mode test.
Cool Down
Skin
Redo BG
Later
I Calibration failed because
your skin temperature is
too high.
Cool the skin at the
insertion site. Perform
another Blood Glucose
Mode test. If calibration
fails 3 times, remove the
Sensor and insert a new
one (see page 43).
* Urgency is defined as: L = Low, I = Intermediate, M = Medium, H = High