User's Manual

Clinical Trial Support Line: 1-888-971-4766 - 157 -
ART05005-REV-4
Alarm
Message
Urgency* What It Means What To Do
Projected
Low Glucose
M This is an early warning
that your glucose level
and the rate at which it is
changing indicate that you
may soon reach your Low
Glucose Threshold.
When set to High
sensitivity, you probably
have about 30 minutes
before reaching your Low
Glucose Threshold.
When set to Medium
sensitivity, you probably
have about 20 minutes
before reaching your Low
Glucose Threshold.
When set to Low
sensitivity, you probably
have about 10 minutes
before reaching your Low
Glucose Threshold.
Follow the
recommendations of your
healthcare team for
treating or stopping the
drop in glucose that you
are experiencing.
Receiver-
related
Alarms
Alarms
Inactive
Replace
Receiver
I Receiver malfunction.
Problem has persisted for
1 hr.
Call the Clinical Trial
Support.
Receiver
Error
Alarms
Inactive
I Receiver error. Wait 50 minutes for the
problem to go away. If
problem persists, call
Clinical Trial Support.
Log Cleared L History log has been
successfully cleared.
Acknowledge the alarm.
* Urgency is defined as: L = Low, I = Intermediate, M = Medium, H = High