Alarm
Message
Urgency* What It Means What To Do
Log
Corrupted
Some Data
Lost
L History log is corrupt. The Receiver detected a
corrupt sector in the log
and erased it. The data in
this sector is lost. Your
reports will not contain this
lost data. If the error occurs
again, call Clinical Trial
Support.
Preferences
Lost
Reconfigure
L An error occurred with the
Receiver configuration.
The Receiver has reset.
Your stored preferences
may be lost.
Review your alarm
settings.
Replace
Receiver
Battery Within
1 Week
L The batteries in your
Receiver will lose power in
approximately 1 week.
Replace the Receiver
batteries (see page
28)before you insert a new
Sensor. Then reset the
time and date (see page
30). Replacing the
batteries before a Sensor
has expired will require you
to insert a new Sensor.
Replace TX
Battery With
Next Sensor
L The Transmitter battery
monitor indicates there are
only 4 days of battery
power left.
Replace the Transmitter
battery prior to inserting the
next Sensor. Be sure to
reconnect the Transmitter
and Receiver.
TX Battery
Exhausted
Replace
Battery
L Transmitter battery power
loss is imminent.
Remove Sensor, replace
Transmitter battery,
reconnect Transmitter and
Receiver, and insert a new
Sensor.
Sensor-
related
Alarms
Alarms
Inactive
Insert New
Sensor
I Alarms are not operating
because the 3-day life of
the Sensor has ended.
Remove the Sensor and
insert a new one (see page
43).
* Urgency is defined as: L = Low, I = Intermediate, M = Medium, H = High