User's Manual

Clinical Trial Support Line: 1-888-971-4766 - 164 -
ART05005 REV 4
Error Code What It Means What To Do
Er3
Incorrect procedure (for
example, blood was
applied to Test Strip
before Test Strip was
inserted into Receiver).
Problem with the Test
Strip.
Problem with the
Receiver.
Be sure you see Apply Sample
on the Receiver screen before you
apply blood or Control Solution.
Perform a Control Solution test
using a new Test Strip. If results
are within the range printed on
Test Strip vial, retest your blood
using a new Test Strip.
If Control Solution Test does not
work or error code persists, call
Clinical Trial Support.
Er4
Problem with Test
Strip.
Problem with
Receiver.
Perform a Control Solution Test
using a new Test Strip. If results
are within the range printed on
Test Strip vial, retest your blood
using a new Test Strip.
If Control Solution Test does not
work or error code persists, call
Clinical Trial Support.
Er5 N/A This error code not currently in use.
Er6 Button on Receiver was
pressed while you were
applying blood or Control
Solution.
Retest using a new Test Strip. Do NOT
touch the Right Option Button while
applying blood or Control Solution to the
Test Strip. If error code persists, call
Clinical Trial Support.