Clarke Error Grid Analysis
Table 1 shows the distribution of the System/YSI data pairs within the zones of the
Clarke Error Grid. Overall, 98.2% of the pairs were within the clinically accurate A
Zone or the clinically acceptable B Zone of the error grid.
Table B.1: Clarke Error Grid Analysis of Data from FreeStyle Navigator
TM
System Accuracy Study
Clarke Error Grid
Zone
Number of Pairs Percent Cumulative
Percent
A 7865 70.7 70.7
B 3056 27.5 98.2
C 36 0.3 98.5
D 165 1.5 100.0
E 0 0.0 100.0
Interfererents
In vitro and in vivo testing suggests that usual pharmacologic levels of ascorbic acid
have no effect on the function of the System but salicylic acid has minimal effect. In
vitro testing suggests that normal physiologic levels of uric acid, lipids, and bilirubin
do not affect System function. The impact of oral hypoglycemic agents and other
potential interfering substances has not been studied.
Limitations
Inaccuracy in glucose values obtained using the BG mode during calibration of the
System will affect the accuracy of the System.