User's Manual

Clinical Trial Support Line: 1-888-971-4766 - 178 -
ART05005 REV 4
Receiver: 12 seconds after last user action
Receiver memory
60 days of normal use including continuous
glucose readings (stored every 10 minutes) and
daily blood glucose readings
Date/time will be remembered for 10 minutes
after Receiver battery removal
Calibration
Plasma equivalent
Calibration time
First Calibration: This should be done approximately
one hour after a new Sensor has been inserted.
There is no maximum time allowed for this. Data
CANNOT begin to be calculated until this first
calibration is done.
Second Calibration: This must be done between 2
and 4 hours after the first calibration or glucose will
no longer be calculated. The second calibration can
still be performed after 4 hours, and glucose
calculation will then resume.
Third Calibration: This must be done between 21 and
29 hours after the second calibration or glucose will
no longer be calculated. The third calibration can still
be performed after 29 hours, and glucose calculation
will then resume.
Blood glucose test time Average of 7 seconds
Blood glucose assay method Coulometric electrochemical sensor
Blood sample type Whole blood, capillary
Hematocrit 15% to 65%.
Classification
The Transmitter and Receiver are classified as follows: