When your Receiver is dropped, damaged, or exposed to liquids.
When your healthcare team advises you to do so.
CAUTION: Do NOT use FreeStyle Control Solution for calibration.
FreeStyle Control Solution is good for 3 months after opening the bottle, or until the
expiration date printed on the label, whichever comes first. Count forward 3 months
from the date you open a new bottle of Control Solution. This is your discard date.
Write this date on the side of the Control Solution bottle. For example, if you open
the Control Solution on January 15
th
, count forward 3 months to April 15
th
. This is the
discard date that you record on the bottle.
Ensuring accurate Control Solution results:
Use only FreeStyle Control Solution with your Test Strips. The use of other
brands of Control Solution may produce inaccurate results.
Replace the cap on the Control Solution bottle immediately after using.
Do NOT use Control Solution past the expiration date or you may get inaccurate
results.
Do NOT add water or any liquid to Control Solution.
Control Solution tests are specified to be accurate only when performed between
59° and 104°F (15° and 40°C).
CAUTION: Results from Control Solution tests do NOT reflect your blood glucose
level.
The Control Solution test results should fall within the range printed on the Test Strip
vial label. If Control Solution test results are out of this range, repeat the test. Out-
of-range Control Solution test results may be caused by:
Expired or bad Control Solution.
Expired or bad Test Strips.