Error in performing test.
Watered-down Control Solution.
Code on Test Strip vial does not match code set in Receiver.
Receiver malfunction.
Control Solution test done outside 59° to 104°F (15° to 40°C).
IMPORTANT: The Control Solution range is a target range for Control Solution only.
It is NOT a target range for your blood glucose levels.
If your Control Solution test results continue to fall outside the range printed on the
Test Strip vial, the Receiver may not be working properly. Do NOT use the System
to test your glucose levels. Call the Clinical Trial Support Line at 1-888-971-4766.
Important Test Strip Information:
The System has an operating range of 40 to 104 F (5 to 40 C).
Use the Test Strips only within the System operating temperature range.
Store the Test Strip package in a cool, dry place between 37 to 86F (3 to 30 C).
Keep Test Strip package away from direct sunlight and heat.
Store your Test Strips in their original vial only; never transfer them to another
vial or any other container or carrying case.
Never store individual Test Strips outside the vial.
After removing a Test Strip from the vial, immediately replace the vial cap and
close tightly.
With clean, dry hands, you may gently touch the Test Strip anywhere when
removing it from the vial or inserting it into the Receiver.
Do NOT use Test Strips beyond the expiration date printed on the package as
this may cause inaccurate results.
Do not bend, cut, or alter Test Strips in any way.
For detailed storage and usage information, refer to the FreeStyle Test Strip
package insert.