User's Manual

Clinical Trial Support Line: 1-888-971-4766 - 36 -
ART05005 REV 4
NOTE: If you do not change or set the Test Strip code number within 3 seconds, the
Receiver automatically moves on to the next step and will accept a Control Solution
using the existing Test Strip code number. If you need to change the Test Strip code
number, press the Left Option Button to go Back to the previous screen, or simply
remove the Test Strip and reinsert it. Then carefully select the correct Test Strip
code number as described in Step 2.
3. The Receiver will prompt you to Apply Sample (see below).
4. Gently touch only one edge of the Test Strip to the Control Solution. Specifically,
apply the Control Solution to the Test Strip edge next to one of the dark-colored
half-circles. If Progress Tones are turned on (see page 111), the Receiver will
beep once to indicate that enough Control Solution was applied.
IMPORTANT: Do NOT apply Control Solution to both edges of the Test Strip.