User's Manual

Clinical Trial Support Line: 1-888-971-4766 - 68 -
ART05005 REV 4
3. Once the Test Strip is inserted, the Receiver will automatically display the most
recently used Test Strip code number. Compare the code number on the Screen
to the code number on the Test Strip vial. If the code numbers match, press the
Right Option Button to select Set. If the code numbers do not match, use the
Up/Down Arrow Buttons to change the code, and, once the code on the screen
matches the code on the Test Strip vial, press the Right Option Button to select
Set.
NOTE: If you do not change or set the Test Strip code number within 3 seconds, the
Receiver will accept a blood sample using the existing Test Strip code number and
automatically move on to the next step. If you need to change the Test Strip code
number, press the Left Option Button to go Back to the previous screen, or simply
remove the Test Strip and reinsert it. Then carefully select the correct Test Strip
code number as described in Step 3.
NOTE: To ensure accurate test results, MAKE SURE the code number on the
Receiver matches the code number on the Test Strip vial.