User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-40
Continuous Burst Phaco Power (Panel Only)
Continuous Burst delivers a 110 ms ultrasonic burst duration. As you press the
footpedal through Position 3, the bursts get closer together. At the maximum level
of footpedal Position 3, the bursts blend together, and the power becomes
continuous (at the preset power level).
Override PHACO Submode Settings
You can override the settings for a PHACO submode by selecting the submode,
and either press the Up and Down arrows to increase or decrease Aspiration Rate,
Vacuum, or Power settings, or press Settings on the control panels for Aspiration
Rate, Vacuum, or Power. If you pressed the Settings button, the Settings pop-up
window appears.
Figure 5.32 – Override PHACO Submode Settings