User's Manual

5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-44
OCCLUSION MODE™ Phaco Settings
You can set different Aspiration Rates for occluded aspiration as opposed to
unoccluded aspiration.
With the OCCLUSION MODE™ Phaco you can set a different vacuum rise time
when the phaco tip is occluded without changing the aspiration rate through an
unoccluded needle.
To set the Occluded Aspiration Rate thresholds:
1. Press PHACO.
2. Press on the left of the screen. Three new control panels appear for
Aspiration and Power. The center panel is for a graphical display of the CASE
settings.
3. Press On to set the thresholds.
4. Change the settings in the new control panels as needed.
5. Press Finished to close the window.
Figure 5.36 – OCCLUSION MODE™ Phaco
6. Press Finished to close the screen.