5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-47
OCCLUSION MODE™ Phaco Power Settings
You can set different power levels when the phaco tip is occluded.
When you are in OCCLUSION MODE™ Phaco, there is an additional control
panel for Power.
Figure 5.39 – OCCLUSION MODE™ Phaco Power Settings
To adjust the occluded power delivery:
1. In OCCLUSION MODE™ Phaco, press the Up or Down arrows to increase or
decrease the occluded power level.
2. Press Settings to change other power settings. A Power Settings screen opens.
3. Press Linear or Panel. Depending on whether you select Linear or Panel,
there are a total of nine different power settings you can choose.
4. Press the buttons on the right of the Settings screen to select a power delivery
value.
5. Press the On button to engage WHITESTAR
®
Technology.
6. Press Finished to close the screen.
Venting an Occlusion
When the aspiration port is blocked or occluded by tissue or other material, the
vacuum pressure builds up. The aspiration flow system vents to the bottle when the
footpedal is released. Another choice is that you can release the footpedal to
Position 1 and that causes the aspiration system fluid to vent using pump rotation.
1. WHITESTAR
®
Technology with Pulse Shaping with Ellips™ Technology
2. WHITESTAR
®
Technology without Pulse Shaping