5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 5-48
These methods release the material at the aspiration port and gives you full control
if the tip accidentally grabs the capsule or iris. The internal fluidic system allows
the desired vacuum level to be maintained when you hold the footpedal at a
constant position. The two adjustments associated with aspiration flow are Max
Vac and Max Flow.
Passive Reflux You can select from either Yes or No. The Yes reverses the pump to vent. The No
option opens a path to the bottle pressure and allows reflux into the eye.
1. Select Configuration.
2. Select Surgeons and Programs.
3. Select the applicable operating mode and submode.
4. Press the Passive Reflux button to select either Yes or No. The system default
is Yes.
Ellips™ Technology Press the Ellips™ Technology button on the Phaco Power Settings screen to
activate. The is shown on the Phaco operating mode screen when Ellips™
Technology is on. Ellips™ Technology can be set for each Phaco submode.
Note: You must have an Ellips™ Phaco handpiece attached to the system before
you can activate the Ellips™ Technology.
Figure 5.40 – Phaco with Ellips™ Technology