User's Manual

10 • Error Messages Troubleshooting and Diagnostics
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 10-2
Error Message
Display
The red Error Messages are shown at the top of the screen. Press the Help button to
open the corrective action for that error. When the error is corrected, press OK to
clear the error message from the screen. If the error message is cleared but not
corrected, the error message is generated again.
The yellow Alerts are shown at the top of the screen. The alert does not need to be
cleared as an error message. An alert, for example, can be: Remote Control Battery
power level is 10 – 25% of maximum. The battery needs to be charged.
WARNING: DO NOT try to replace the Wireless Remote Control
batteries. Call your AMO Technical Service representative to replace the batteries.
Fuse Replacement
Procedure
If the System does not turn on when the power switch is turned On, and you have
confirmed that the power cord is properly connected and plugged in, check to see if
the fuse is bad.
Note: To prevent the risk of fire or damage to the instrument, replace the fuses
with the exact type and rating as indicated below (check the voltage sticker
on the back panel of System to confirm your system voltage):
Table 10.1 – Fuse Specifications
To replace the console fuses:
1. Unplug the System electrical power.
2. Unplug the power cord from the back panel.
3. Locate the fuse holder on the back panel of the System, as shown below.
4. Use a small screwdriver to gently pry open the cover and expose the fuse
holder.
5. Gently pry out the fuse holder.
6. Remove the bad fuse and replace the fuse with a new fuse (value and size
specified above).
7. Replace the fuse holder. Make sure that the arrows point to the right side of the
back panel. Tilt the fuse holder slightly to the right and push in.
8. Push the fuse holder cover up and in until the cover snaps closed.
9. Reconnect the power cord to the back panel.
10. Plug the System into an electrical receptacle (outlet).
!
Voltage Quantity Fuse
Specifications
Console 100/120/240 2 6.3A, 250V,
Bussman GDA