10 • Error Messages Troubleshooting and Diagnostics
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 10-6
17 FP Pitch Parameter Error Hardware failure or
program error.
Select Next Case.
18 IV Check Error Hardware failure or
program error.
Select Next Case.
19 Vit Cutter Check Error Hardware failure or
program error.
Select Next Case.
101 Fluidics communication
error.
Invalid data from the IH
or communication errors.
Select Next Case.
102 Fluidics write error. Microcontroller SPI, SPI
Bus, EEPROM, or ADC
is faulty.
Cycle power.
103 Fluidics read error. Microcontroller SPI, SPI
Bus, EEPROM, or ADC
is faulty.
Cycle power.
104 Fluidics Reservoir Vacuum
Level Error.
External air pressure low.
Venturi muffler clogged.
External air valve
clogged.
External air leak.
Venturi vacuum path leak.
Vacuum sensor not
calibrated or faulty.
Reprime the System.
105 Fluidics Reservoir Leak. Venturi vacuum path leak Reprime the System.
106 Fluidics Tank Vacuum
Level Error.
Reservoir vacuum is low.
Tank vacuum path leak.
Vacuum sensor not
calibrated or faulty.
Vacuum regulator not
calibrated or faulty.
Vacuum regulator drive
circuit failure.
Reprime the System.
107 Fluidics Tank Leak. Tank vacuum path leak. Reprime the System.
Error
Number
Message Probable Cause Corrective Action