User's Manual

10 • Error Messages Troubleshooting and Diagnostics
WHITESTAR SIGNATURE™ System Rx Only – Z370147 Rev. C 1108 10-7
108 Fluidics Pack Vacuum
Level Error.
Reservoir vacuum is low.
Pack vacuum path leak.
Vacuum sensor not
calibrated or faulty.
Vacuum regulator not
calibrated or faulty.
Vacuum regulator drive
circuit failure.
Reprime the System.
109 Fluidics Pack Leak. Pack leak. Reprime the System.
110 Fluidics RAM error. Bad Microcontroller. Cycle power.
111 Fluidics ROM error. Bad Microcontroller. Cycle power.
112 Fluidics Master
Communication error.
Instrument Host
malfunction.
Bad microcontroller.
Cycle power.
113 Drain Pump Fault. Droplets, bubbles, or fog
on chamber walls.
Bad fluid level sensor(s).
Bad cable.
Fluid level sensor drive
circuits faulty.
Drain pump drive circuit
fault.
Check the chamber and the
bag for over-inflation.
Select Next Case.
114 Fluidics DAC Fault. Circuit failure (I2C bus,
DAC, or micro).
Reprime the System.
115 Fluidics Evacuation
Vacuum Level Error.
Reservoir vacuum is low
Evacuation vacuum path
leak
Vacuum sensor not
calibrated or faulty
Vacuum regulator not
calibrated or faulty
Vacuum regulator drive
circuit failure
Reprime the System.
116 Fluidics Evacuation Leak. Evacuation path leak. Reprime the System.
Error
Number
Message Probable Cause Corrective Action